Abstract

A simple HPLC method for determination of Benzalkonium chloride in various ophthalmic solutions was developed. The chromatographic analysis was achieved using CN column (250 mm, 4.6 mm i.d., 5 μm) with isocratic mobile phase consisting of acetonitrile - phosphate buffer (pH 5.5; 0.05 M) (70:30, v/v) at a flow rate of 1 mL/minute. The column temperature was maintained at 25°C and the detection wavelength was 210 nm.
The proposed HPLC method was successfully validated according to the ICH guideline and proved to be stability-indicating.
This method was applied to quantify Benzalkonium chloride during in-use stability study of two ophthalmic solutions. Antimicrobial effectiveness of Benzalkonium chloride in these solutions was also evaluated.
The developed method is suitable for the routine analysis of Benzalkonium chloride in many ophthalmic solutions as well as for the stability studies.

Key words: Benzalkonium chloride, Ophthalmic solutions, HPLC, Stability-indicating, In-use stability.






Full-text options


Share this Article


Online Article Submission
• ejmanager.com




ejPort - eJManager.com
Refer & Earn
JournalList
About BiblioMed
License Information
Terms & Conditions
Privacy Policy
Contact Us

The articles in Bibliomed are open access articles licensed under Creative Commons Attribution 4.0 International License (CC BY), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.