Improving animal-specific radiotherapy quality assurance for kilovoltage X-ray radiotherapy using a 3D printed dog skull water phantom

Yoshinori Tanabe, Toshie Iseri, Ryouta Onizuka, Takayuki Ishida, Hidetoshi Eto, Munekazu Nakaichi.




Abstract
Cited by 0 Articles

Background:
Accurate dose assessment during animal radiotherapy is beneficial for veterinary medicine and medical education.
Aim:
To visualize the radiation treatment distribution of orthovoltage X-ray equipment in clinical practice using Monte Carlo simulations and created a dog skull water phantom for animal-specific radiotherapy.
Methods:
EGSnrc-based BEAMnrc and DOSXYZnrc codes were used to simulate orthovoltage dose distributions. At 10, 20, 30, 40, 50 and 80 mm in a water phantom, depth dose was measured with waterproof Farmer dosimetry chambers and the diagonal off-axis ratio was measured with Gafchromic EBT3 film to simulate orthovoltage dose distributions. Energy differences between orthovoltage and linear accelerated radiotherapy were assessed with a heterogeneous bone and tissue virtual phantom. The animal-specific phantom for radiotherapy quality assurance was created from CT scans of a dog and printed with a three-dimensional printer using polyamide 12 nylon, with insertion points for dosimetry chambers and Gafchromic EBT3 film.
Results:
Monte Carlo simulated and measured dose distributions differed by no more than 2.0% along the central axis up to a depth of 80 mm. The anode heel effect occurred in shallow areas. The orthovoltage radiotherapy percentage depth dose in bone was >40%. Build-up was >40%, with build-down after bone exit, whereas linear accelerator radiotherapy absorption changed little in the bone. A highly water-impermeable, animal-specific dog skull water phantom could be created to evaluate dose distribution.
Conclusion:
Animal-specific water phantoms and Monte Carlo simulated pre-treatment radiotherapy is useful quality assurance for orthovoltage radiotherapy and yields a visually familiar phantom that will be useful for veterinary medical education.

Key words: Orthovoltage, Animal radiotherapy, Heel effect, Water animal phantom, Animal specific radiotherapy quality assurance






Full-text options


Share this Article


Online Article Submission
• ejmanager.com




ejPort - eJManager.com
Refer & Earn
JournalList
About BiblioMed
License Information
Terms & Conditions
Privacy Policy
Contact Us

The articles in Bibliomed are open access articles licensed under Creative Commons Attribution 4.0 International License (CC BY), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.